Villa Sistemi Medicali
ARCOVIS 3000 S/R
User's Manual
[File:201175-G-01-01.doc]
Rev. B - Pag. 39/76
Utility Mode - Report
The report screen is a particular screen accessible to the user in only reading. It does not
include the data necessary to the normal unit use.
Firmware Version
Indication of the firmware versions loaded on the unit.
Serial Number
Indication of the monobloc and x-ray tube serial number.
Tot. Time Xray – min.
Counter of the total time (expressed in minutes) of the exposures in
fluoroscopy mode with relative resetting key.
Tot. Rad Esposure
Counter of the total exposures number in Radiography mode with
relative resetting key.
Install date
Unit installation data with relative resetting key. Datum inserted at the
moment of the unit installation.
Maintenance Date
Limit date foreseen for the unit maintenance with relative resetting key.
Datum automatically calculated by the unit according to the installation
date or the date of the last maintenance (foreseen yearly).
Key of resetting/activation. The activation/resetting keys are not
accessible to the final user. For the possible data resetting it is
necessary the intervention of Service.
Key to exit from the display.