User's Manual
ARCOVIS 3000 S/R
Villa Sistemi Medicali
Pag. 70/76 - Rev. E
[File:201175-G-01-01.doc]
5.8. Accessories
Description
Standard
Optional
X-ray handswitch with extensible cable
•
Fluoroscopy control by triple footswitch
•
Cassette-holder for radiography on cassette 24x30cm
•
Cassette-holder for radiography on cassette 18x24cm
•
Cassette-holder for radiography on cassette 10x12inch
•
Sterile drapes set
•
Thermal dry film printer Sony UP 970 / UP 990 or equivalent
•
Medical Image Capture Device MediCap USB200
•
USB Data: 512 MB, 350 Kb per image, about 1450 image
•
USB Data: 2 GB, 350 Kb per image, about 5700 image
•
DVD Recorder Sony DVO 1000 MD. Automatic start- stop with DIP3000
•
DVD Recorder for Fluoro acquisition, Only manual Star Stop
•
Dicom Utility ESIDIC2 (only for DIP3000 memory)
•
Dicom Utility ESIDIC3 (only for HRC memory)
•
Laser Targeting device (only on I.I. tube)
•
Dose meter with ionization chamber (DAP) Diamentor CM-T
•
Printer for dose meter S Sprint s (on unit)
•
Collimator with iris and parallel shutter
•
Equipotential node cable lg.300
•
External interlocks kit
•
Movement handles on “C” arm
•
Movement handle on I.I. tube
•
Laser targeting device, for dose reduction (only on I.I. tube)
Description
Data
Wave length
635nm
Optical power of the laser diode
< 4 mW
Optical power of the collimated beam
< 3,5 mW
Divergence
6,67 mRad
Optical power on the operation plane
< 3 mW
Safety class
1A
Dose meter with ionizing chamber PTW, DIAMENTOR CM-T ( (DAPMeter)
Description
Data
Measurement amplifier
Measurement category
Dose area product
Measuring unit
μ
Gy m
2
Digital resolution
0,01
μ
Gy m
2
Measurement category
Measurement category
Dose area product rate
Measuring unit
μ
Gy m
2
/s
Digital resolution
0,01
μ
Gy m
2
/s
Nominal useful range
Dose rate range
0,10 to 400 mGy/sec
Max field dimension
φ
6 cm
Min field dimension
φ
0,6 cm
Dose area product range rate
0,01 to 4500
μ
Gy m
2
/s
Tube voltage
40 to 120 kV
Chamber:
Type TA34037
Dimension of chamber field
φ
7,2 cm
Max energy dependence
±
8% according to IEC60580, Table 6
Hardening equivalent value
< 0,3 mmAl / 70 kV according to IEC60522/1999