2.7. Operative messages
Display in fluoroscopy mode
1 Indication of the dose value (only if the dosimeter is present and in working condition /
MESSAGES and/or ERRORS Area)
2
Indication of the FLUOROSCOPY / RADIOGRAPHY phase.
3 Indication of the available thermal units (HUT), expressed in percentage from 0% (indicator
completely white and spent HUT) to 100% (indicator completely black and available HUT).
When the residual HUT value allows a use of the unit below five minutes, under the
indicator will appear also the residual time indication (t = …min), valued in function of the
employed power. Simultaneously it’s issued an acoustic alarm not modifiable.
When the residual HUT reach 0%, the x-rays emission block intervenes.
It’s possible to exceed this limit and proceed with the x-rays emission (only in fluoroscopy
mode) till intervention of monobloc thermal security.
This choice must be performed by Service personnel and on express request of the final
user.
As the monobloc temperature decreases, the HUT will be regenerated. Residual time
indication will disappear, when thermal units allow a use exceeding five minutes.
4 Selection of the Anatomic Curve in Fluoroscopy (Standard / Paediatric / Pelvis-Head /
Standard for
sturdy
patient
(
only for rotating anode with camera 1K
2
)
).
5 Indication of the set kV value. It changes in Automatic mode or in Manual mode by using
the kV+ and kV- keys of the control panel.
6 Indication of mA value concerning kV and the selected curve. During the exposure it
indicates the measured mA value.
7
Indication of the AUTOMATIC / MANUAL mode.
8
Dose printing key. It appears if the dosimeter is present and in working condition and if the
printing mode in Utility Mode is selected (§3.6.10).
1
2
3
4
5
6
7
8