Villa Sistemi Medicali
ARCOVIS 3000 S/R
User's Manual
[File:201175-G-01-01.doc]
Rev. A - Pag. 3/76
1.1. Electrical safety
Only trained service personnel authorized by Villa Sistemi Medicali may
remove the unit covers and only in accordance with the instructions
contained in the Service Manual.
This x-ray unit may only be used in environments or medical rooms in compliance with the
applicable IEC standards.
The x-ray unit must not be used in areas where there exists a danger of explosion.
Cleaning and disinfecting agents, including those used on patients, may create an explosive,
gaseous mixture. Use only those products in compliance with the applicable rules.
1.2. Laser targeting devices safety
•
Keep always a good lightening on the room.
•
Never look through the output window of the laser targeting device.
•
Never fix the reflections of the laser targeting devices.
•
Before starting any examination, the patient must remove earrings, glasses, necklaces
and whatever could reflect the laser beam and be printed on the image.
•
Don't clean the openings of the laser targeting devices with tools that could modify their
optics. Only the service personnel must perform possible cleaning actions.
•
The min. distance between the laser source and the patient must not be lower than cm20.
The only purpose of the laser use is to reduce the patient dose to a
minimum, it doesn’t be considered as an absolute means of centering.
The activation of procedures other those listed above can cause the
emission of dangerous non-ionizing radiations.
1.3. Mechanical safety
The unit can be moved only by trained personnel by complying with the following safety
conditions:
•
Move the unit by considering the conditions for the movement that are indicated in detail
in the paragraph 3.1.
•
Don't move and use the unit on surfaces with inclination higher than 5°
•
Only use the proper handles to move the unit.
•
Avoid collision with obstacles.
•
After positioning the unit, engage the parking brakes.