User's Manual
ARCOVIS 3000 S/R
Villa Sistemi Medicali
Pag. 48/76 - Rev. F
[File:201175-G-01-01.doc]
5.3. Electrical data
Description
Fixed Anode Version
Rotating Anode Version
Voltage
230 Vac ±10% standard monophase
105 / 115 / 125 / 220 / 240 Vac ±10% monophase on request
Frequency
50 Hz standard - 60 Hz on request
Absorbed current in fluoroscopy
4,5A @ 230Vac
7,5A @ 115Vac
5A @ 230Vac
10A @ 115Vac
Absorbed current in radiography
20A @ 230Vac
28A @ 115Vac
20A @ 230Vac
30A @ 115Vac
Absorbed current in stand-by
1,0A @ 230Vac
1,7A @ 115Vac
1,0A @ 230Vac
2,0A @ 115Vac
Line compensation
Automatic
Line resistance
<0,4
Ω
@230V
<0,2
Ω
@115V
Standard mains plug
16A @230Vac
Classification EN60601-1 par.5:
Protection against electrical hazards
Class I
Protection against direct and indirect contacts
equipment with applied part Type B
Protection against water seepage
common protection (IPXO)
Use conditions
Continuous operating with intermittent load
The unit is not suitable to be used where there exist mixtures inflammable with air or nitrous oxide
5.3.1. RADIOLOGICAL DATA
Description
Fixed Anode Version
Rotating Anode Version
Generator, power in DC current
(IEC 60601-2-7)
2,5kW @100kV
(100kV, 25mA, 100ms)
3,5kW (100kV, 35mA, 91ms)
mAs < 2,2 @230Vac;
5kW (100kV, 50mA, 64ms)
mAs < 2,2 @230Vac;
Generator, max power in DC current
2,5kW (100kV, 25mA, 128ms)
mAs < 2,2 @115Vac)
3,5kW (100kV, 35mA, 91ms)
mAs < 2,2 @115Vac
Max high voltage (fluoroscopy and
radiography)
110kVp 120kVp
Frequenza Inverter
40 kHz
Max. current in continuous fluoroscopy
8,0mA
(STANDARD curve)
Max. current in “SNAPSHOT” fluoroscopy
10mA
12mA (0,5K camera)
30mA (1K camera)
Max. current in radiography
25mA @230Vac
18mA @115Vac
Max. mAs in radiography
125mAs @230Vac
90mAs @115Vac
Description
Fixed anode version
Rotating anode
version
Max. fluoroscopy time at 75W (75kV-1mA)
continuous continuous
Max. fluoroscopy time at 280W (80kV-3,5A)
44 min
54 min
Max. fluoroscopy time at 400W (100kV-4mA)
29 min
38 min
Max. fluoroscopy time at 525W (70kV-7,5mA)
24 min
31 min
Max. fluoroscopy time at 550W (110kV-5mA)
21 min
---
Max. fluoroscopy time at 600W (120kV-5mA)
--- 28
min
Max. fluoroscopy time
H.U Safety after 21 min
of fluoroscopy @110kV,
5mA (550W).
H.U Safety after 28 min
of fluoroscopy @120kV,
5mA (600W).
Cooling at 30°
300 min
420 min