Villa Sistemi Medicali
ARCOVIS 3000 S/R
User's Manual
[File:201175-G-01-01.doc]
Rev. A - Pag. 65/76
5.7. Digital image processor
TV chain and Monitor with CCD 0,5K x 0,5 K
Features
SBFM 76
SBFM 78/330
SBFM 78/2700
Model
SBFM76
SBFM 78/330
SBFM 78/2700
Images number
L.I.H.
Only Ram image
L.I.H. +330
(non volatile images)
L.I.H. + 2700
(non volatile images)
Image format
768x576x12 bits 50Hz
256 gray level
768x576x12 bit 50Hz
256 grey levels
768x576x12 bit 50Hz
256 grey levels
A/D Converter
8 bits
10 bit
10 bit
D/A Converter
8 bits
8 bit
8 bit
Sampling frequency
15 MHZ
15 MHZ
15 MHZ
Monitors number
1 2 2
Recursive Filter, noises
reduction, OFF,2,4,8,16
YES
0,2,4,8,16
YES
0,2,4,8,16
YES
0,2,4,8,16
Digital rotation in real time
YES
YES YES
Vertical image Inversion
YES, in combination
with the rotation
YES, in combination
with the rotation
YES, in combination
with the rotation
Horizontal image inversion
YES
YES YES
Grey scale Inversion
NO YES YES
Frame rate fps acquisition
Single store image
Single store image
Single store image
Image edge (EDGE)
YES YES YES
Patient data editing
NO YES YES
Video input
Standard CCIR 1 Vpp
composite video signal
75 Ohm termination
Composite Video
signal
standard CCIR 1Vpp
end 75 Ohm
Composite Video
signal
standard CCIR 1Vpp
end 75 Ohm
Video output
BNC, Standard CCIR 1
Vpp composite video
signal
75 Ohm termination
BNC, Composite
Video signal
standard CCIR 1Vpp
end 75 Ohm
BNC, Composite
Video signal
standard CCIR 1Vpp
end 75 Ohm
Power supply
20V to 36V dc
10Vdc
÷
36Vdc
10Vdc
÷
36Vdc
Dimensions
BOX (26x10,2x6,5)
BOX (26x10,2x6,5)
BOX (26x10,2x6,5)
Weight
....
...