USER'S MANUAL
Operating instructions
(Rev. 1)
ROTOGRAPH-D (120V)
21
6.2.2
6.2.2
6.2.2
6.2.2
Programmed/Manual exposure
Programmed/Manual exposure
Programmed/Manual exposure
Programmed/Manual exposure
After setting the unit as previously described previous point, it is now
possible to choose whether to operate an exposure in a preset mode,
meanings with the kV values preset by the manufacturer or to operate
an examination in the manual function, with possibility to change the
preset kV values.
6.2.2.1
6.2.2.1
6.2.2.1
6.2.2.1
Programmed exposure
Programmed exposure
Programmed exposure
Programmed exposure (refer to Figure 8)
Push button "33" for adult program or "34" for child's.
To check the new configuration depress push buttons Adult (33) and
Child (34), there will only be a variation on kV and time; pressing
buttons Large Size (35), Medium Size (36) and Small Size (37), there will
be a variation on kV.
The following table shows the programmed values according to the
selection made:
Large Size
(35)
Medium Size
(36)
Small Size
(37)
Adult
(33)
80kV - 17sec
70kV - 17sec
65kV - 17sec
Child
(34)
65kV - 14sec
60kV - 14sec
60kV - 14sec