USER'S MANUAL
Operating instructions
ROTOGRAPH-D (120V)
(Rev. 1)
22
6.2.2.2
6.2.2.2
6.2.2.2
6.2.2.2
Manual exposure
Manual exposure
Manual exposure
Manual exposure (refer to Figure 8)
To select the manually technical factors follow the next procedure:
1.
Depress button kV (38), the button function will blink.
2.
Depress buttons increase (41) or decrease (40) on the hand control
selecting the desired kV value to be used. After 2 seconds, the
function button LED stops blinking. The unit is now ready for the
next step.
NOTE:
The kV can vary from a minimum of 60kV to a maximum of 85kV with
steps of 5kV.
If, after having reached the minimum or maximum values of the kV
range additional decrease or increase is requested, the following
messages will be displayed:
Minimum
or
Maximum
; the minimum or
maximum set kV values will subsequently be displayed.