Defects due to wrong setting of radiological data and dark
Defects due to wrong setting of radiological data and dark
Defects due to wrong setting of radiological data and dark
Defects due to wrong setting of radiological data and dark
room
room
room
room
•
Under
Under
Under
Under----exposed or over
exposed or over
exposed or over
exposed or over----exposed image
exposed image
exposed image
exposed image
The number of kV set are not adequate for the patient's build, more
likely an error in development.
Corrective action:
Repeat the examination with a different value: see paragraph 6.2.4. If
the error happens again, call the
Technical Assistance
•
Completely white image
Completely white image
Completely white image
Completely white image
It has not been exposed to X-rays or digital sensor borken.
Corrective action:
Verify the X-rays emission by checking light and acoustic signal
(light 43 Figure 8), If the defect persists call for
Technical
Assistance
.