80
Pic 76
electric meters log screen [Archives>Electricity]
8.6
Discrete inputs log [Archives>Values>Discrete inputs]
Discrete input table contains records about discrete channel changes. Totally data logger support 8
discrete channels.
Every record has time stamp, channel number (from 0 to 7, totally -8), state after changes (0 –
open, 1 – closed contact) and state of all channels, in binary format you can see state of every
channel (0 –open, 1 – closed contact), if value is two symbols, this means it is in HEX (you need
to convert to binary manually) (see
Pic 77
screen of discrete inputs log [Archives>Discrete imputes]
8.7
Event logging [Archives>Values>Events]
Event conditions:
•
Analog inputs value go out from set range,
•
Discrete input channel change.
Every deviation can be set/changed by configuring data logger. On deviation appears, it will be