6. Use and Warranty Period
Do not attempt to disassemble the device without authorization. The warranty period
is 2 years from the date of purchase. For shelf-life of accessories please refer to the
appropriate manual. During the warranty period, the warranty will be void under the
following conditions:
1) An error caused by operating the unit in any unprescribed conditions or applications.
2) Damage caused by using an improper power supply, improper installation or
operations not listed in this manual.
3) Damage caused by installation, modification (the technical description is included in
the user manual) or repair from unauthorized service engineers.
4) Damage caused by natural disasters such as fire, earthquake, power surge, lightning,
flood, etc.
When the machine develops a fault, please contact your supplier or Vincent Medical for
maintenance.
Customer Service Contact Information
Vincent Medical Manufacturing Co., Ltd.
Flat/RM B2, 7/F., Hang Fung Industrial Building, Phase 2, 2G Hok Yuen Street, Hung Hom,
Kowloon, Hong Kong
Tel: +852 2186 1010
E-mail: service@inspired-medical.com
Website: www.inspired-medical.com
Obelis s.a
Bd Général Wahis, 53, 1030 Brussels, Belgium
Tel: +32 2 732 59 54
Fax: +32 2 732 60 03
E-Mail: mail@obelis.net
5.4 Recommended Separation Distances Between Portable
and Mobile RF Communications Equipment and the Model
VUN-001
Recommended Separation Distances Between
Portable and Mobile RF Communications Equipment and the Model VUN-001
150kHz to 80MHz
d=1,2×P
80MHz to 800MHz
d=1,2×P
800MHz to 2,5GHz
d=2,3×P
The model VUN-001 is intended for use in an electromagnetic environment in which
radiated RF disturbances is controlled. The customer or the user of the model VUN-001
can help to prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the
model VUN-001 is recommended below, according to the maximum output power of
the communications equipment.
Rated maximum
output power of
transmitter w
0,01
0,1
1
10
100
0,12
0,38
1,2
3,8
12
0,12
0,38
1,2
3,8
12
0,23
0,73
2,3
7,3
23
Separation distance according to frequency of transmitter m
For transmitters rated at a maximum output power not listed above, the recommend-
ed separation distance in meters (m) can be estimated using the equation applica-
ble to the frequency of the transmitter, where is the maximum output power rating
of the transmitter in watts (W) accordable to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz. the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
½
½
½
d
P
43