2. inspired™ O2FLO Parts, Features
and Specification
The inspired™ O2FLO is intended to deliver warmed and humidified respiratory gases
to spontaneously breathing patients with a flow rate from 2LPM to 60LPM and is
capable of providing high flow therapy. The device can be used in the emergency
room, outpatient department, inpatient department and other diagnostic and
treatment rooms. This device can also be used for homecare.
a. Dew Point Temperature Control
This device allows the operator to set the dew point temperature from 31°C to 37°C and
is equipped with highly sensitive temperature probes to monitor patient end tempera-
ture. The device also has a number of temperature protection settings to ensure patient
safety.
b. Oxygen Concentration Control
This device can be used with an additional oxygen supply. The device accurately
displays real-time oxygen concentration of gas mixture delivered. The user can control
the oxygen concentration by adjusting the external oxygen valve. The upper and lower
limits of oxygen concentration alarm is settable to ensure patient safety.
c. Flow Rate Control
This device allows the operator to set the flow rate from 2LPM to 60LPM. The device
monitors the flow rate with a built-in highly sensitive flow rate sensor.
d. Mode Control
The device has two operating modes, namely Adult Mode and Paediatric Mode. The
device will identify if the correct breathing circuit is being used (either Adult or Paediat-
ric) to avoid misuse.
e. Display and Control Interface
The device is equipped with a LCD display monitor and control knob to make the
device easier to control, and to provide clear display of functions and data.
f. Alarms
The device has an auditory alarm function to alert patient or end user when an alarm
condition occurs. The auditory alarm is accompanied by indicator light and information
displayed on the monitor.
1.1 Intended Use
2.1 inspired™ O2FLO Main Unit
1.2 Product Features
① ② ③
④
⑤
⑥
⑦
⑧
⑩
⑨
11
12 13
14
18
21
19
15
16
17
20
7
8