Front Panel
Rear Panel
Sensor Head, Waveguide and Taper
Display
reads power on
selected range
Range Switch
Select ranges from 200 µW
to 200
mW. ‘Remote’
enables USB control and
auto ranging
Sensor Head
Connector Port
with standard Amphenol
cable connected
Zero Button
Press to re-zero display
Calibration
Factor Switch
(dB)
Used to adjust the power
meter display
Internal
Calibration
Switch
On/Off
Toggle switch
Fuse
0.25A, SB
AC Power Input
(90-240V, 50-60 Hz)
Calibration Input
Banana jacks for DC
calibration (max 15V
)
USB Interface
Type B jack
Analog Output
via BNC to other
equipment; -10 to +10
VDC
Front Panel / Rear Panel / Sensor Head Overview
Section
Two
Sensor Head
with a 1-inch WR10 straight waveguide
and a 1-inch waveguide taper to WR10
are attached.
Copper
Flange
is nickel plated;
tapped holes are
easily stripped. Keep
waveguide in place as
shown; do not over-
tighten screws.
1-Inch Waveguide Taper
Many standard tapers are available from
VDI (not included with meter; may be
purchased separately) for measurements
in bands above WR10.
1-Inch WR10 Straight
Waveguide
attached to protect the Flange.