flowscreen
Operating Instructions
12 / 24
Version 1.0
Copyright ©
3.4.3
Defining sensors and connections
• <
Settings
> <
Sensor X Selection / Calibration
>
Select Channel
Pin allocation in the sensor cable plug (see 3.11). Factory setting is
Pin 2
.
3.4.4
Calibrating the sensors
• <
Settings
> <
Sensor X Selection / Calibration
>
Calibration
The sensors are set ex-works with the values 0 bar (
Zero Value
) and 16 bar (
Full Scale
).
If other sensors are to be connected, these values must be adjusted according to the
manufacturer's information.
3.5
Base settings
3.5.1
Results / error messages
• <
Settings
> <Results>
<
Error after x faults
> = After this number of consecutive defective dosings (pressure value
outside of measuring range), the error LED illuminates red.
3.5.2
Start and end of the measurement
• <
Settings
> <
Sensor X Selection / Calibration
>
Measurement Start/End at
Selection:
Pressure
Enter the pressure value that defines
Start/End
Start/Stop-Signal
Signal analysis (system plug (4))