7.4.2 Recording Group
Recording group commands are only valid while the system is in the idle state.
7.4.2.1 Manual Recording Mode
Length
Command
Group
Command
Code
Card Mask
Channel
Mask
Parameters
0x0007
0x02
0x0001
Used
Ignored
0x00
Byte 8:
Recording Mode
0x00 Off
0x01 Single
Shot
0x02 Continuous
This command sets the manual recording mode for the selected cards. There are three available manual
recording modes:
Off - disables manual recording (default)
Single Shot - records one reading for each channel
Continuous - records continuously until manual recording is disabled
7.4.2.2 Set Pre-trigger Buffer Size for Manual Recording Command
Length
Command
Group
Command
Code
Card Mask
Channel
Mask
Parameters
0x000A
0x02
0x000D
Used
Ignored
0x00
Bytes 8-11:
32-bit buffer size, in scans.
Sets the “pre-trigger” buffer size for manual recording
Range: 0 to 645,000
Default: 0