Technical Manual
5-16
Flow rate update time
The flow rate value is updated as shown in the table below. At start-up
the accuracy of the flow rate will not gain full precision until after a full
derivation time period.
Number of weight conversions per
derivation time period
Flow rate value update interval
<= 100
Every weight conversion
> 100 but <= 200
Every second weight conversion
> 200 but <= 300
Every third weight conversion
And so on . . .
And so on . . .
Flow rate value to analog output
The flow rate value can be obtained on a connected analog output.
The settings for analog outputs are found in the menu 'Analog outputs'
(chapter ‘Set-up’ in this manual).
Supervision of the flow rate value
It is possible to level supervise the flow rate value and use it in the setpoint function.
The settings for level supervision and setpoint functions are found in the menus 'Level
1' to 'Level 32' and ‘Setpoints’ (chapter ‘Set-up’ in this manual).
Communication
The flow rate value can be read from a register by using communication.
It’s also possible to control if flow rate or weight should be shown on the display.