Technical Manual
7-6
To resolve such a blocking situation without being able to access the blocking interface
it is possible to restart the instrument using the ‘Instrument Restart’ function that can be
found in the ‘Maintenance’ menu. This function must be used with caution since it will
interrupt all instrument functions while restarting the instrument. Also note that if the
‘Start Mode’ is set to ‘Command’ the instrument will wait at start-up for a start
command from the instrument front panel or via communication.
There are a few differences between local and remote access. From the Remote
Access Interface it is not possible to do Program Upgrade (Maintenance) or Network
Configuration (Main Menu). All other functionality is available.
Remote Set-up
All operating functions in the G4 Instrument are controlled by parameters. The
parameter values are permanently stored in the instrument and will not be lost when
the unit is switched off. At delivery the parameters are factory-set to default values,
giving the instrument an initial standard function.
The actual setting of the parameter values can be read and edited during normal
measuring operation in sub menu ‘Parameter Set-up’. When Scale Batching (a
program option) is active, on any scale, it will not be possible to edit parameters except
for a limited number of batching parameters that still can be edited.
Editing of parameter values can be performed from a browser when connected and
logged in to the instrument. After editing hardware parameters the instrument will be
restarted.
In the instrument there are two levels of security locks provided to protect from
unauthorized access to instrument functions and editing of parameters and values. The
locks are opened by four-digit codes.
Warning: Changes done during editing of set-up parameters will affect the behavior of
the instrument immediately. The user must take all necessary precautions to prevent
any undesired effects in the process monitored or controlled by the G4 instrument or a
connected control system.
It is strongly recommended to activate the set-up lock in the instrument to prevent any
unauthorized changes of set-up parameters.
It’s a good practice to make a backup of the set-up after changes have been done. See
section ‘Maintenance’ in this chapter for more information on backup and restore.
When the G4 Instrument is first taken into service, setting of a few parameters will
quickly adapt the instrument to the transducers and give desired weighing function. The
‘Quick Set-up’ is described in a separate publication:
G4 Multi Channel Weighing Instrument
Operating instructions, Quick installation
RM Type
Summary of Contents for G4
Page 2: ......
Page 18: ...Technical Manual 1 14 ...
Page 39: ...G4 Multi Channel Weighing Instrument 3 7 ...
Page 104: ...Technical Manual 5 18 ...
Page 158: ...Technical Manual 6 54 ...
Page 178: ...Technical Manual 7 20 ...
Page 186: ...Technical Manual 8 8 ...
Page 196: ......
Page 197: ...G4 Multi Channel Weighing Instrument Appendix 1 ...
Page 198: ...Technical Manual ...
Page 199: ......