G4 Multi Channel Weighing Instrument
3-27
Range/Alternatives
Explanation and
<default value>
result of alternatives.
1:T.Signal P1
Range: +/–9.99999
Unit: mV/V
<0.00000>
In this parameter, enter the recorded value for the
transducer signal in the first calibration point.
1:T.Signal P2
Range: +/–9.99999
Unit: mV/V
<1.66631>
In this parameter, enter the recorded value for the
transducer signal in the second calibration point.
1:T.Signal P3
Range: +/–9.99999
Unit: mV/V
<2.03900>
In this parameter, enter the recorded value for the
transducer signal in the third calibration point.
1:T.Signal P4
Range: +/–9.99999
Unit: mV/V
<2.03900>
In this parameter, enter the recorded value for the
transducer signal in the fourth calibration point.
1:T.Signal P5
Range: +/–9.99999
Unit: mV/V
<2.03900>
In this parameter, enter the recorded value for the
transducer signal in the fifth calibration point.
1:T.Signal P6
Range: +/–9.99999
Unit: mV/V
<2.03900>
In this parameter, enter the recorded value for the
transducer signal in the sixth calibration point.
1:Set Zero
See ‘1:Set zero ‘ under section ‘Data sheet calibration
related parameters’ for explanation of the parameter.
1:Zero Offset
See ‘1:Zero offset ‘ under section ‘Data sheet
calibration related parameters’ for explanation of the
parameter.
Gross Weight:
This help display of live gross weight for the current
scale gives the possibility to check the load on the
scale at any moment.
Transd. Signal:
This help display of live transducer signal for the
current scale gives the possibility to check the signal at
any moment.
Summary of Contents for G4
Page 2: ......
Page 18: ...Technical Manual 1 14 ...
Page 39: ...G4 Multi Channel Weighing Instrument 3 7 ...
Page 104: ...Technical Manual 5 18 ...
Page 158: ...Technical Manual 6 54 ...
Page 178: ...Technical Manual 7 20 ...
Page 186: ...Technical Manual 8 8 ...
Page 196: ......
Page 197: ...G4 Multi Channel Weighing Instrument Appendix 1 ...
Page 198: ...Technical Manual ...
Page 199: ......