G4 Multi Channel Weighing Instrument
5-5
Gross/Net operation
At normal operation the instrument presents a numerical weight value at the display,
either gross weight or net weight. When net weight is displayed the text ‘N’ is
displayed on the lower line.
Toggling between display of gross weight and net weight can be performed as
explained under section ‘Weight display’. Note that toggling is done only for the
selected scale (shown on the display).
Net weight is the difference between gross weight and a tare value. For calculation of
net weight, the instrument can use either ‘Preset’ tare, ‘Auto’ tare, or the sum of them.
Net weight cannot be displayed if the tare value in use is zero (0).
Zero setting
A basic zero setting of the gross weight is performed as part of the calibration for the
scale. If changes to the weighing equipment are made later a renewed calibration, or at
least the zero setting of a calibration, should be performed.
Minor correction of the zero value may be needed and can rapidly be performed:
When a gross weight close to zero is displayed, go to the zeroing, see section
‘Weight display’. This will make the gross weight zero.
Pressing the zeroing key sequence, see section ‘Weight display’, will also set the value
of ‘Auto tare’ to zero.
Zero setting with the zeroing key sequence is permitted only if:
•
The weight is stable (the letter ‘M’ not shown) and
•
The accumulated zero correction since last calibration is between -1 % and +3 %
of the ‘Capacity’, over and above the zero offset obtained when the instrument
was calibrated last time.
The text ‘Z’ will be shown on the display when the displayed weight is a ’good zero’,
meaning that the weight deviates from zero with less than one quarter of the set
‘Resolution’.
Zero value
Unloaded transducers.
Actual transducer load.
Tare value: auto,
fixed or auto + fixed
Net weight
Gross weight
Range for zero adjustment
and zero-tracking.
Empty scale - loaded with
permanent equipment only.
Installation setting.
Correlation between gross weight, net weight,
and tare value for a scale.
Summary of Contents for G4
Page 2: ......
Page 18: ...Technical Manual 1 14 ...
Page 39: ...G4 Multi Channel Weighing Instrument 3 7 ...
Page 104: ...Technical Manual 5 18 ...
Page 158: ...Technical Manual 6 54 ...
Page 178: ...Technical Manual 7 20 ...
Page 186: ...Technical Manual 8 8 ...
Page 196: ......
Page 197: ...G4 Multi Channel Weighing Instrument Appendix 1 ...
Page 198: ...Technical Manual ...
Page 199: ......