BewellConnect - BW-PFX1 - User Manual - 072017v1
BewellConnect - BW-PFX1 - User Manual - 072017v1
68
69
EMC data
1. This product needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided, and this unit can be af-
fected by portable and mobile RF communications equipment.
2. Do not use a mobile phone or other devices that emit electromagnetic fields, near the
unit. This may result in incorrect operation of the unit.
3. Caution: This unit has been thoroughly tested and inspected to assure proper perfor-
mance and operation!
4. Caution: this machine should not be used adjacent to or stacked with other equipment
and that if adjacent or stacked use is necessary, this machine should be observed to
verify normal operation in the configuration in which it will be used.
Refer to further guidance below regarding the EMC environment in which the device
should be used.
GUIDANCE AND MANUFACTURE’S DECLARATION – ELECTROMAGNETIC EMISSION
The BW-PFX1 is intended for use in the electromagnetic environment specified below. The customer of
the user of the BW-PFX1 should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The BW-PFX1 use RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF emission
CISPR 11
Class B
The BW-PFX1 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
GUIDANCE AND MANUFACTURE’S DECLARATION – ELECTROMAGNETIC IMMUNITY
The BW-PFX1 is intended for use in the electromagnetic environment specified below. The customer or
the user of BW-PFX1 should assure that it is used in such an environment.
Immunity test
EN 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or ceramic
tile. If floor are covered with synthetic material,
the relative humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2kV for power
supply lines
Main power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
±1 kV differen-
tial mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips,
short interrup-
tions and volt-
age variations on
power supply
input lines IEC
61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the BW-PFX1 requires continued operation
during power mains interruptions, it is recom-
mended that the BW-PFX1 be powered from an
uninterruptible power supply or a battery.
Power frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
3 A/m
30 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Summary of Contents for BewellConnect BW-PFX1
Page 3: ...BewellConnect BW PFX1 User Manual 072017v1 4 5 FR MyPeriTens BW PFX1 0197...
Page 15: ...BewellConnect BW PFX1 User Manual 072017v1 28 29 NL MyPeriTens BW PFX1 0197...
Page 34: ...BewellConnect BW PFX1 User Manual 072017v1 67 E M C D ATA...
Page 37: ...BewellConnect BW PFX1 User Manual 072017v1 73 WA R R A N T Y...