screen. The analysis is above the measurement range selected,
either superior to 42.9°C (109°F) in
Body
Mode or superior to
60°C (140°F) in
Surface Temp
Mode.
- The screen displays the message LO: when using the
Ther-
moFlash® LX-26 EVOLUTION
, the message
Lo
can show on the
screen. The temperature analysed is under the measuring range
selected, either less than 32°C (90°F) in
Body
Mode or less than
0°C (32°F) in
Surface Temp
Mode.
This message displays in various cases – please find below a list
of the main cases:
Reasons for
LO
message display
Advice
Temperature reading hampered by
hair, perspiration…
Make sure that there is no
obstruction prior to taking
a temperature.
Temperature hampered by an air flux. Make sure there is no air flux as this
could interfere with the infrared system.
Temperature readings too close to-
gether, the
ThermoFlash®
did not
have the chance to boot itself.
Respect the pause of 15 seconds
minimum between two readings –
1 minute pause is advised.
The measuring distance is too far.
Please respect the measuring dis-
tance (between 5 and 8 cm – 2 in
and 3.14 in).