background image

STRUCTURE           

The Structure of 

MRP-5000 

Microscope

 OPERATING 

INSTRUCTIONS

  

Set light dimmer control to lowest intensity and switch on the power, increasing slowly 
to desired level. Begin by placing a microscope slide containing a specimen to be 
observed on the object stage and clamp it carefully with the movable spring clip. Adjust 
the interpupillary distance between the two eyepieces to the most comfortable position 
for the user. The left eyepiece is equipped with a diopter adjustment to fit the user’s 
needs.   

Using the X and Y Direction Knobs located just below the stage (on the right hand side) 
the specimen may be moved to the center of the stage for observation.  Once the 
specimen is in focus, the X and Y Direction Knobs may be adjusted to observe different 
sections of the specimen.   

Begin observation of your specimen with the 4X Objective. Turn the coarse focusing 
knob until a clear image is obtained, then use the fine focus knob to enhance the 
observation of the specimen to its clearest image. On the coarse focus knob is a tension 
control ring (located on the right side of the microscope against the stand) which can be 
turned to tighten or loosen the tension of the focusing knob to the user’s desired setting. 

When the desired view is obtained under the lowest power (4X), rotate the nosepiece to 
the next higher magnification (10X). The nosepiece should “click” into position. It 
should only be necessary to adjust the fine focus knob slightly to again have a clear view 
of the specimen. As the magnification is increased, the user may also find adjustments in 
the various methods of regulating the brightness (see below) will improve the view of 
the specimen.  

It is important to remember when adjusting the focus that the objective should never 
touch the specimen. When increasing to 100X magnification, the objective will appear to 
be very close to the slide. Because the 100X is an oil immersion objective, a drop of 
cedar-wood oil or other immersion oil should be applied in the gap between the 
objective and the specimen. (Note:  all slides should be prepared with cover slips over 
the specimen for observation under the 100X objective.) The 100X oil immersion 
objective should be wiped off with a piece of soft clean cloth or lens tissue to remove the 
immersion oil immediately after using. 

This microscope is equipped with a “locking nut” (located behind the stage against the 
arm of the stand) to help prevent the stage from moving up far enough to collide with the 
objectives. This locking nut is pre-set in position at the factory and should not require 
adjustment. If you find that the stage is not moving up far enough to get your specimen 
in focus, try turning the locking knob slightly to allow the stage to travel farther. 

1.

 

Eyepiece 

2.

 

Graduated Diopter Ring 

3.

 

Nosepiece with Objectives 

4.

 

Objective Stage 

5.

 

Adjustable Locking Nut 

6.

 

Tension Control for Focusing Knob 

7.

 

Coarse Focus Adjustment Knob 

8.

 

Fine Focus Adjustment Knob 

9.

 

Mechanical Stage X-Y Adjustment Knob 

10.

 

Condenser with Iris Diaphragm and Filter Holder 

11.

 

Field Diaphragm for Kohler Illumination 

12.

 

Dimmer Control for Brightness Adjustment 

 

12

11 

 3 

 

6

 4 

 

9

 

7

 

 

8

 

5

10 

Reviews: