background image

www.vitalbaby.com

Made in China to Vital Baby

®

 specifications. Vital Innovations Limited, Hatfield, 

Herts, AL10 9EW, United Kingdom. Vital Baby

®

 is a registered trademark.  

PROTECT™ is a trademark. Design Rights.  © Vital Innovations Ltd 2020

4 in 1 contactless 

thermometer

PROTECT

PROTECT

PROTECT

PROTECT

Summary of Contents for PROTECT HTD8216C

Page 1: ...4 in 1 contactless thermometer PROTECT PROTECT PROTECT PROTECT...

Page 2: ...k professional advice immediately when there is a temperature elevation and if you are taking temperature on o babies under 3 months consult your physician immediately if the temperature exceeds 37 4...

Page 3: ...vity or the use of a compress prior to taking the reading If engaged in physical activity and or used any kind of forehead compress wait at least 30 minutes before taking your temperature Temperature...

Page 4: ...ure results may vary if positioned in the wrong location or if used differently to the user instructions set out in this user guide For patients measuring their own temperature it is recommended to us...

Page 5: ...high fever between 38 1 C 42 9 C or 100 6 F 109 2 F This is a possible high fever If the temperature taken indicates a slight or high fever the thermometer will make 3 short beeps when displaying the...

Page 6: ...memory are indicated by an M at the bottom of the display STEP 3 The thermometer will then beep once and the screen will dim this means it has entered standby mode During standby mode you can change...

Page 7: ...hen be taken Hold the thermometer close to the forehead approx 1cm and press and hold the start button while simultaneously scanning the forehead side to side See Figure 4 You will hear 6 long beeps f...

Page 8: ...thermometer will perform a start up test during which the screen will light up green and all display features will show for approx 2 seconds See figure 2 STEP 2 After the start up test is complete the...

Page 9: ...ndicated in the battery compartment If the thermometer is not used for long periods of time remove the batteries to avoid damage Always dispose of used batteries at appropriate collection sites accord...

Page 10: ...to the transmitter manufacturer NOTE At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE These guidelines may not apply in all situations Electromagnetic propaga...

Page 11: ...ronment 10 Vm 80 MHz to 2700 MHz Not applicable Not applicable Not applicable 30A m Field strengths outside the shielded location from fixed RF transmitters as determined by an electromagnetic site su...

Page 12: ...ace the thermometer in the room that you intend to take the reading and allow to acclimatise for 20 minutes before taking another reading Environmental temperature may be unstable Place the thermomete...

Page 13: ...0 4 F 42 1 C 42 9 C 0 3 C 107 8 F 109 2 F 0 5 F Surface mode 2 C 3 6 F Display 4 digits plus special icons resolution 0 1 C 0 1 F Auto power off 35s Measuring time 5 seconds in ear mode 3 10s forehead...

Page 14: ...ational or local regulations Complies to RoHS Restriction of Hazardous Substances guidelines Complies to the WEE directive Do not dispose of this product in the household waste at the end of its usefu...

Reviews: