VITALAB MICRO & MICROLAB 200 THEORY OF OPERATION
VITAL SCIENTIFIC SERVICE MANUAL
2.17
Figure 2.7 The Flowcell unit
2.5 SWITCHED POWER SUPPLY.
The switched power supply is mounted underneath the processor
board on the base-plate.
The AC line voltage is routed via a linefilter to the switched
power supply. The power supply accepts input voltages from 110
to 240 Volts with an frequency from 47 to 63 Hz. Because of
this design it is not necessary to switch over from 110VAC to
220VAC. The primary fuses of 2,5AT are located inside the
powerswitch. The power supply has a primary fuse of 2 A Medium
Time Lag on board.
CAUTION : Serious safety hazard exists on the metal parts of
the switched power supply because of high voltage.
Remove the power cable before working on the unit.
2.5.1 input/output connections.
There is one AC-input and one DC-output connector. On the
8-pole output TB2 the following pins are used:
pin 1 +5VDC ; processor board (red)
pin 3 Ground ; processor board (black)
pin 4 Ground ; blower (blue)
pin 5 +12VDC ; blower (red)
pin 6 +12VDC ; processor board (green)
To the 2-pole input TB1 AC power is supplied:
pin 1 line (brown)
pin 2 neutral (blue)
Note 1:
The ground connection of the board is made by one of
the mounting screws (close to C6). To prevent hazard
or instable readings always check this connection
after disassembling this unit.
Note 2:
The ground connection of the line entry (with switch)
Summary of Contents for Vitalab Micro
Page 1: ...VITALAB MICRO MICROLAB 200 VITAL SCIENTIFIC SERVICE MANUAL VITAL SCIENTIFIC SERVICE MANUAL...
Page 3: ...VITALAB MICRO MICROLAB 200 VITAL SCIENTIFIC SERVICE MANUAL...
Page 4: ......
Page 8: ...VITALAB MICRO MICROLAB 200 VITAL SCIENTIFIC SERVICE MANUAL iv...
Page 15: ...VITALAB MICRO MICROLAB 200 PREFACE VITAL SCIENTIFIC SERVICE MANUAL xi...
Page 16: ......
Page 28: ...VITALAB MICRO MICROLAB 200 GENERAL DESCRIPTION VITAL SCIENTIFIC SERVICE MANUAL 1 12...
Page 54: ......
Page 77: ...VITALAB MICRO MICROLAB 200 INSTALLATION MAINTENANCE VITAL SCIENTIFIC SERVICE MANUAL 3 23...
Page 92: ...VITALAB MICRO MICROLAB 200 INSTALLATION MAINTENANCE VITAL SCIENTIFIC SERVICE MANUAL 4 8 board...
Page 95: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 3 exploded cover...
Page 97: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 5 exploded baseplate...
Page 99: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 7 exploded optical block...
Page 106: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 14 6001 616 1 5 CPU...
Page 107: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 15...
Page 108: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 16 6001 616 2 5 KIO I O...
Page 109: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 17...
Page 111: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 19...
Page 113: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 21...
Page 114: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 22 6001 616 5 5 Drivers...
Page 115: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 23...
Page 117: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 25...
Page 119: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 27...
Page 121: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 29...
Page 123: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 31...
Page 124: ......
Page 126: ...MICRO MICROLAB 200 SERVICE INFO S VITAL SCIENTIFIC SERVICE MANUAL 6 2 11 12 13 14 15...