VITALAB MICRO & MICROLAB 200 INSTALLATION & MAINTENANCE
VITAL SCIENTIFIC SERVICE MANUAL
3.17
3.
The lamp holder can be removed now by pulling it upwards.
4.
Pull out the lamp (it can be HOT!) and replace it by a new
one.
Note:
Never touch the lamp with the bare fingers, as this
contaminates the bulb. Clean the lamp with alcohol if
necessary.
5.
Place the lamp holder back and fasten screw A.
6.
Switch the instrument on and do lamp adjustment.
3.5.11 Lamp adjustment.
1.
Loosen screw B a little and adjust screw A for maximum
transmission (height adjustment).
2.
Adjust screw C also for maximum transmission (horizontal
adjustment).
3.
Fasten screw B, place the top cover etc.
3.5.12 Valve replacement.
The valve unit is mounted with 2 screws. When replacing the
valve unit, take care to place the rings under the special
screws. Do not use other screws than indicated, because this
can cause malfunction of the valve. Note that the solenoid of
the valve does not function without the tubings installed.
Sometimes and especially with new tubings it can happen that
the pinched tube sticks together. Normally it will open due to
the pressure of the pump, if not stretch and roll the tube
through your fingers.
3.5.13 Bellows replacement.
Empty the fluid system first, by means of the flush key. Switch
off the instrument and remove the cover.
The complete pump unit can be removed by loosening the four
Summary of Contents for Vitalab Micro
Page 1: ...VITALAB MICRO MICROLAB 200 VITAL SCIENTIFIC SERVICE MANUAL VITAL SCIENTIFIC SERVICE MANUAL...
Page 3: ...VITALAB MICRO MICROLAB 200 VITAL SCIENTIFIC SERVICE MANUAL...
Page 4: ......
Page 8: ...VITALAB MICRO MICROLAB 200 VITAL SCIENTIFIC SERVICE MANUAL iv...
Page 15: ...VITALAB MICRO MICROLAB 200 PREFACE VITAL SCIENTIFIC SERVICE MANUAL xi...
Page 16: ......
Page 28: ...VITALAB MICRO MICROLAB 200 GENERAL DESCRIPTION VITAL SCIENTIFIC SERVICE MANUAL 1 12...
Page 54: ......
Page 77: ...VITALAB MICRO MICROLAB 200 INSTALLATION MAINTENANCE VITAL SCIENTIFIC SERVICE MANUAL 3 23...
Page 92: ...VITALAB MICRO MICROLAB 200 INSTALLATION MAINTENANCE VITAL SCIENTIFIC SERVICE MANUAL 4 8 board...
Page 95: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 3 exploded cover...
Page 97: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 5 exploded baseplate...
Page 99: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 7 exploded optical block...
Page 106: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 14 6001 616 1 5 CPU...
Page 107: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 15...
Page 108: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 16 6001 616 2 5 KIO I O...
Page 109: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 17...
Page 111: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 19...
Page 113: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 21...
Page 114: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 22 6001 616 5 5 Drivers...
Page 115: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 23...
Page 117: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 25...
Page 119: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 27...
Page 121: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 29...
Page 123: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 31...
Page 124: ......
Page 126: ...MICRO MICROLAB 200 SERVICE INFO S VITAL SCIENTIFIC SERVICE MANUAL 6 2 11 12 13 14 15...