VITALAB MICRO & MICROLAB 200 INSTALLATION & MAINTENANCE
VITAL SCIENTIFIC SERVICE MANUAL
4.3
Symptom A:
Insufficient sample aspiration in any mode. Can
also be caused by an obstruction.
Cause 1:
Leakage of fluid system.
Action:
Check the sipper tube for correct
tightening. Do not over tighten, as
excessive force may either destroy the seal
or strip the threads on the fitting. Check
the seals of the sipper tube connections.
Check all tubing of the fluid system, and
also flowcell and pump.
Remark:
Air is aspirated instead of sample.
Unstable display readings is an additional
indication.
Symptom B:
Insufficient aspiration, but flush mode operates
normally.
Cause 1:
Programm error.
Action:
Check sipper volume in test parameters.
4.3 Temperature error message.
Symptom A:
The message 'TEMPERATURE ERROR' is displayed on
the LCD display. This is an indication that the
instrument has switched off the temperature
control circuit. This will occur when the
selected temperature has not been reached within
20 seconds!
Cause 1:
Ambient temperature too high, >35 C
Cause 2:
Fan not running.
Action:
Check fan, replace if necessary.
Cause 3:
Loose or wrong connected connectors from Peltier
elements or temperature sensor.
Action:
Check the connectors for good connection
and position.
Summary of Contents for Vitalab Micro
Page 1: ...VITALAB MICRO MICROLAB 200 VITAL SCIENTIFIC SERVICE MANUAL VITAL SCIENTIFIC SERVICE MANUAL...
Page 3: ...VITALAB MICRO MICROLAB 200 VITAL SCIENTIFIC SERVICE MANUAL...
Page 4: ......
Page 8: ...VITALAB MICRO MICROLAB 200 VITAL SCIENTIFIC SERVICE MANUAL iv...
Page 15: ...VITALAB MICRO MICROLAB 200 PREFACE VITAL SCIENTIFIC SERVICE MANUAL xi...
Page 16: ......
Page 28: ...VITALAB MICRO MICROLAB 200 GENERAL DESCRIPTION VITAL SCIENTIFIC SERVICE MANUAL 1 12...
Page 54: ......
Page 77: ...VITALAB MICRO MICROLAB 200 INSTALLATION MAINTENANCE VITAL SCIENTIFIC SERVICE MANUAL 3 23...
Page 92: ...VITALAB MICRO MICROLAB 200 INSTALLATION MAINTENANCE VITAL SCIENTIFIC SERVICE MANUAL 4 8 board...
Page 95: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 3 exploded cover...
Page 97: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 5 exploded baseplate...
Page 99: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 7 exploded optical block...
Page 106: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 14 6001 616 1 5 CPU...
Page 107: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 15...
Page 108: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 16 6001 616 2 5 KIO I O...
Page 109: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 17...
Page 111: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 19...
Page 113: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 21...
Page 114: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 22 6001 616 5 5 Drivers...
Page 115: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 23...
Page 117: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 25...
Page 119: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 27...
Page 121: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 29...
Page 123: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 31...
Page 124: ......
Page 126: ...MICRO MICROLAB 200 SERVICE INFO S VITAL SCIENTIFIC SERVICE MANUAL 6 2 11 12 13 14 15...