VITALAB MICRO & MICROLAB 200 INSTALLATION & MAINTENANCE
VITAL SCIENTIFIC SERVICE MANUAL
4.5
4.6 Inconsistent measurement results.
Inconsistent measurement results can be defined as follows:
drop-outs, bad reproducibility, and too low measurement
results.
4.6.1 Drop outs.
Symptom A:
Occasional drop-outs in the measurement. Either
too high or too low.
Cause 1:
Air-bubbles in flowcell as a result of pollution
of the fluid system.
Action:
Flush the system with a solution of
re-distilled water and 5% to 10% neutral
detergent for at least 10 minutes by 37 C.
Cause 2:
Air-bubbles in flowcell due to leakage of the
fluid system.
Action:
Check all tubing connections of the fluid
system, especially those before the
flowcell.
Cause 3:
Air-bubbles in the flowcell as a result of too
cold sample/reagent mixture.
Action:
Problem can be solved by adding a little
neutral detergent to the reagent (0.1%).
Remark:
At low temperatures the viscosity of liquid
increases rapidly. The result will be an
increase of the vacuum inside the fluid
system and air-bubbles may be formed
spontaneously.
4.6.2 Bad reproducibility.
Symptom A:
Measurement results are not reproducible. There
are no actual drop-outs but the variation in the
results (controls) is too high.
Summary of Contents for Vitalab Micro
Page 1: ...VITALAB MICRO MICROLAB 200 VITAL SCIENTIFIC SERVICE MANUAL VITAL SCIENTIFIC SERVICE MANUAL...
Page 3: ...VITALAB MICRO MICROLAB 200 VITAL SCIENTIFIC SERVICE MANUAL...
Page 4: ......
Page 8: ...VITALAB MICRO MICROLAB 200 VITAL SCIENTIFIC SERVICE MANUAL iv...
Page 15: ...VITALAB MICRO MICROLAB 200 PREFACE VITAL SCIENTIFIC SERVICE MANUAL xi...
Page 16: ......
Page 28: ...VITALAB MICRO MICROLAB 200 GENERAL DESCRIPTION VITAL SCIENTIFIC SERVICE MANUAL 1 12...
Page 54: ......
Page 77: ...VITALAB MICRO MICROLAB 200 INSTALLATION MAINTENANCE VITAL SCIENTIFIC SERVICE MANUAL 3 23...
Page 92: ...VITALAB MICRO MICROLAB 200 INSTALLATION MAINTENANCE VITAL SCIENTIFIC SERVICE MANUAL 4 8 board...
Page 95: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 3 exploded cover...
Page 97: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 5 exploded baseplate...
Page 99: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 7 exploded optical block...
Page 106: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 14 6001 616 1 5 CPU...
Page 107: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 15...
Page 108: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 16 6001 616 2 5 KIO I O...
Page 109: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 17...
Page 111: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 19...
Page 113: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 21...
Page 114: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 22 6001 616 5 5 Drivers...
Page 115: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 23...
Page 117: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 25...
Page 119: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 27...
Page 121: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 29...
Page 123: ...MICRO MICROLAB 200 EXPLODED VIEWS VITAL SCIENTIFIC SERVICE MANUAL 5 31...
Page 124: ......
Page 126: ...MICRO MICROLAB 200 SERVICE INFO S VITAL SCIENTIFIC SERVICE MANUAL 6 2 11 12 13 14 15...