3. Check the manometer and the cuff pressure. Pressure will
inflate to 70mmHg. Verify that the pressure is within ±2% or
±3mmHg. If OK click on the
Pass
box to place a checkmark in
the box. Click
Next
.
Figure 6-31: Pressure Accuracy Test, 70 mmHg
4. Repeat for 140 and 280mmHG.
Figure 6-32: Pressure Accuracy Test, 140 mmHg
Figure 6-33: Pressure Accuracy Test, 280 mmHg
5. Click
Done
if the test passes after the 280 mmHg test.
6. Turn off power and remove serial cable and manometer fixture.