Criticare Systems, Inc.
VitalCare
TM
506N3 Series
Service Manual
Page 7-3
Section 7 —Disassembly
Replace the Printer
1. Follow the caution for static-sensitive devices in “Electrostatic
Discharge Protection” in this section.
2. Turn the monitor off and disconnect the monitor from the AC
(Mains) power source.
3. Remove the battery as described in “Battery Removal/
Replacement” in Section 5.
4. With a #1 Phillips screwdriver, remove the two (2) painted
screws (pn 40193B002) from the printer assembly (pn
95745A001) on the back of the monitor.
Figure 7-1: Remove Screws from Printer Assembly
5. Press the front of the printer assembly to release the printer
assembly from the tab.
Figure 7-2: Release Printer Tab
Remove Screws