Section 7 —Disassembly
Page 7-10
VitalCare
TM
506N3 Series
Service Manual
Criticare Systems, Inc.
8. Remove the three (3) screws (pn 40995B005) with nylon
washers (pn 40283B001) that are holding the FasTemp Board to
the side panel.
Figure 7-13: Remove Screws from FasTemp PCB
9. Press the tab at the top of the side panel to release the
FasTemp Board from the side panel.
NOTE: Under the bottom of the Board are two (2) insulating
washers. DO NOT LOSE. These must be positioned as shown.
Figure 7-14: DO NOT Lose Washers
10.Lift the pad that is holding the flex cable to the FasTemp Board.
11.Remove the flex cable from the FasTemp Board by unlocking
the connector (
J3
) of the FasTemp Board.
NOTE
: The silver pads of the flex cable face down to connect to
the FasTemp Board. Same direction pad is placed on the
Isolation Board.
Remove Screws
Washers