Section 7 —Disassembly
Page 7-12
VitalCare
TM
506N3 Series
Service Manual
Criticare Systems, Inc.
Replacing the TurboTemp
Board
Refer to “Replace the Temperature Boards” on page 7-5 for directions
on removing the side panel.
1. Carefully lift the side panel.
2. Remove the Temp Cable (pn 90931A001) from the connector
beneath the PCB (
J1
).
NOTE
: Cable is locking. If the cable is placed into the connector
reversed, the board will not function and may cause damage.
Figure 7-15: Remove Temp Cable and Screws
3. Place side panel with TurboTemp Board facing upwards
(Component side down).
4. Remove the four (4) screws (pn 40995B009) from the
TurboTemp Board (see above). Torque of screws is 3 in. lbs.
5. Carefully lift up. (The side panel has the temp holder attached.
6. Replace the TurboTemp Board.
7. Reassemble in reverse order.
Follow “Completion of Service” procedure at the end of this section.
Remove Screws
Disconnect
Temperature Cable
Torque
@ 3 in lbs