Criticare Systems, Inc.
VitalCare
TM
506N3 Series
Service Manual
Page 7-21
Section 7 —Disassembly
13.Remove the two (2) standoffs (pn 42255B004) and two (2) nuts
(pn 40008B003) that are attached to the carrier PCB. Torque is
1 in. lbs.
14.Replace the Nellcor Carrier PCB (pn 91387A001) and/or
Nellcor PCB (pn 83459B001).
15.Reassemble in reverse order.
Follow “Completion of Service” procedure at the end of this section.