Section 7 —Disassembly
Page 7-28
VitalCare
TM
506N3 Series
Service Manual
Criticare Systems, Inc.
12.Route the black ground wire under the membrane tail
(if necessary).
13.Plug speaker assembly (pn 90932A001) into
J3
of the
Main Board.
14.Slide the speaker into the front rail of the bezel. Place the
orange and brown wires into relief as shown below.
Figure 7-34: Reinsert Speaker
Reassemble monitor and follow “Completion of Service” procedure at
the end of this section.
Front Rail
Relief Strain
Summary of Contents for 506DN3
Page 12: ......
Page 56: ......
Page 62: ......
Page 90: ......
Page 156: ......
Page 184: ......
Page 225: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Page 253: ......
Page 254: ......