Service Manual
Page 7-37
Section 7 —Disassembly
12.Tilt the PCB assembly so that it is on an angle. Lift the bezel up
so that the tail of the membrane slides through the opening of
the PCB assembly. Place the two together and lay back down
onto the table.
13.Mount the PCB assembly (pn 90232A001) to the bezel using
four (4) P.H.M.S. screws (pn 40995B005). (These are the upper
screws. Do Not tighten down.)
14.Unlock connector from membrane and attach the membrane
tail into it. Close the connector by pushing it close. Keep the
membrane straight and even.
15.Fold over the membrane’s ground tail and connect to the Main
Board using a P.H.M.S. screw and washer. Attach the right side
of the bezel and torque all five (5) screws at 5 in. lbs.
NOTE
: If the monitor has an old Main Board (pn 91384A001 or
91384A002), install a P.H.M.S. screw through the ground cable
assemble (pn 95602A001) and secure it to the lower left-hand
corner of the bezel. Attach the ground flex circuit from
membrane by folding over the ground tail and inserting P.H.M.S.
screws through the ring terminal and both holes of the ground
tail. Attach the right side of the bezel and torque all screws at
5 in. lbs. Route the ground wire under the membrane tail.
Figure 7-47: Flex Cable Ground
16.Plug speaker assembly (pn 90932A001) into
J3
of the
Main Board.
Membrane
Ground Tail
Summary of Contents for 506DN3
Page 12: ......
Page 56: ......
Page 62: ......
Page 90: ......
Page 156: ......
Page 184: ......
Page 225: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Page 253: ......
Page 254: ......