Section 1 —Introduction
Page 1-12
VitalCare
TM
506N3 Series Service Manual
Criticare Systems, Inc.
Specifications
DOX SpO
2
Accuracy:
70 to 100: ±2%
50 to 69: ±3%
0 to 49: Unspecified
Statistical, represents one st. dev. (~66%)
of clinical samples
Range:
1-99%
Resolution:
1%
Indications
Numerical, Audible
(pulse tone pitch varies with SpO
2
)
Method:
Dual wavelength LED
Operation:
Continuous Use
Sensor Wavelength:
660nm/905nm
Sensor Power:
<80mW
SpO
2
Pulse Rate Range:
20 to 300
SpO
2
Pulse Rate Accuracy:
±1 bpm
Nellcor SpO
2
(OxiMax)
Accuracy:
70 to 100: ±2%
Below 69: Unspecified
Neonate: 70 to 100: ±3%
Statistical, represents one st. dev. (~66%)
of clinical samples
(Sensor accuracy is found in “Accuracy of
OxiMax Sensors” on page 1-8)
Range:
1-99%
Resolution:
1%
Indications
Numerical, Audible
(pulse tone pitch varies with SpO
2
)
Method:
Dual wavelength LED
Operation:
Continuous Use
Motion Artifact Rejection:
yes
SpO
2
Pulse Rate Range:
20 to 300 bpm
SpO
2
Pulse Rate Accuracy:
±2 bpm (20 to 250 bpm)
Summary of Contents for 506DN3
Page 12: ......
Page 56: ......
Page 62: ......
Page 90: ......
Page 156: ......
Page 184: ......
Page 225: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Page 253: ......
Page 254: ......