Criticare Systems, Inc.
VitalCare
TM
506N3 Series Service Manual
Page 1-15
Section 1 —Introduction
Mechanical/Electrical
Weight:
3.0 kg; 6.7 lbs. (base unit with battery)
Size (without temperature):
22 cm (H) x 18 cm (W) x 16.5 cm (D)
8.7 in. (H) x 7.1 in. (W) x 6.5 in. (D)
Size (with temperature):
22 cm (H) x 22 cm (W) x 16.5 cm (D)
8.7 in. (H) x 8.7 in. (W) x 6.5 in. (D)
Battery:
Rechargeable; Sealed lead acid battery
Rating:
6V, 7.2 Amp Hours
Battery Life:
8 hours, with NIBP every 5 minutes
Recharge Time:
6 hours
Power Requirements:
100 – 240 VAC (±10%), 50 – 60 Hz
Environmental
Operating Temperature:
0° to 45° C (32° to 113°F)
Storage Temperature:
-5° to 50°C (23° to 122°F)
Operating and Storage Humidity:
15% to 90%; non-condensing
Medical Device:
Class II Equipment
Electrical Protection:
Class I Equipment
Degree of Protection:
Type CF, Defibrillator-Proof
Protection against ingress:
IPX1 rating, Drip-Proof Equipment
All specifications are subject to change without notice.
Summary of Contents for 506DN3
Page 12: ......
Page 56: ......
Page 62: ......
Page 90: ......
Page 156: ......
Page 184: ......
Page 225: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Page 253: ......
Page 254: ......