NIBP Calibration Mode
To enter the NIBP Calibration mode:
1. Press the POWER key and the NIBP/START/STAT/STOP key at
the same time.
2. The 506N3 Series monitor attempts to connect to extended
NIBP calibration tools through the external serial port,
identifying itself as a 506N3 Series monitor.
3. The message
CHECKING FOR NIBP TOOLS...
should appear
in the LCD message bar.
A service calibration application, called NIBP SERVICE
(pn 97083A003), may be run on a connected PC. See “NIBP
Calibration” in Section 6 for testing details.