Criticare Systems, Inc.
VitalCare
TM
506N3 Series
Service Manual
Page 2-13
Section 2 —Service Menus
NOTE
: The following is a list of possible menus and submenus.
Depending on the model and configuration of your monitor, some
submenus listed may not be applicable.
Alarm Menu
Press the MENU key to enter menus. Press the UP/DOWN keys until
the
ALARM MENU
displays. Press the MENU key to enter the
ALARM MENU
.
Press the UP/DOWN keys to move through the alarm submenu until
the desired item setting displays in the LCD window.
Press the MENU key to select that item, then use the UP/DOWN keys
to scroll through the setting options for that item. When the desired
setting displays on the LCD, press the MENU key to select that
setting.
Alarm limits are set separately for adult, pediatric, and neonatal
modes and are saved independently.
To set adult alarm limits, enter the
ALARM MENU
while you are in the
adult mode. Change the patient mode by pressing the PATIENT key
until the ADULT LED is lit. Confirm that the ADULT indicator on the
front panel is lit. Set all desired alarm limits for adult monitoring
conditions.
To set pediatric alarm limits, enter the
ALARM MENU
while you are in
the pediatric mode. Change the patient mode by pressing the
PATIENT key until the PEDIATRIC LED is lit. Confirm that the
PEDIATRIC indicator on the front panel is lit. Set all desired alarm
limits for pediatric monitoring conditions.
To set neonate alarm limits, enter the
ALARM MENU
while you are in
the neonate mode. Change the patient mode by pressing the
PATIENT key until the NEONATE LED is lit. Confirm that the
NEONATE indicator on the front panel is lit. Set all desired alarm
limits for neonate monitoring conditions.
Alarm Volume
This setting adjusts the volume of the audible alarm tone. Choices
are 1 through 10 and
Off
(Where 1 is the softest and 10 is the
loudest). A tone is generated at each volume level as the alarm
volume setting is changed.
The factory hospital default setting is 4; the factory alternative care
default setting is 3. The monitor returns to the value 2 upon a power
cycle if the alarm volume had been turned off or set to 1.
Pulse Volume
Sets the volume for the audible pulse tone from
Off
, and 1 through 10,
with 1 being the softest and 10 the loudest. The factory default value
is
Off
.
Adult
Pediatric
Neonate
PATIENT key
LEDs
Summary of Contents for 506DN3
Page 12: ......
Page 56: ......
Page 62: ......
Page 90: ......
Page 156: ......
Page 184: ......
Page 225: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Page 253: ......
Page 254: ......