Section 2 —Service Menus
Page 2-18
VitalCare
TM
506N3 Series
Service Manual
Criticare Systems, Inc.
International
Configuration Settings
The language setting only appears in the
CONFIGURATION
submenu after the monitor has been started in the factory default
mode (Press the MENU key during power up). This setting is
intended to be set once upon arrival at a final destination. The
language setting does not appear again until default settings are
recalled again.
Language
The monitor has language settings available in English, French,
German, Italian, Spanish, Portuguese, Danish, and Dutch. The
monitor must be restarted before the language setting change
activates.
Trend Button Settings
Press and hold the TREND button to set the trend settings.
Interval
Sets the interval of time you wish the data to print in the trend span.
Choices are
1, 2, 5, 10, 15, 30
, or
60 M
(minutes), and
Off
. The
factory default setting is
1 M
.
NOTE
:
Off
is not available if
FORMAT
is set to
NO BPT
.
Format
Sets the format of trended data. Choices are
BP
,
BPT
,
NO BPT
,
LAST
and
ALL.
The factory default setting is
BPT.
Span
Sets the amount of trend data you wish to print. Choices are: 1
0, 15,
30
, or
60 M
(minutes);
2, 4, 8, 12
, or
24 H
(hours); and
ALL
. The
factory default setting is
4 H
.
Patient Button Settings
Press the PATIENT button to select the patient mode. Choices are
Adult
,
Pediatric
, and
Neonatal
. The factory default setting is
Adult
.
Summary of Contents for 506DN3
Page 12: ......
Page 56: ......
Page 62: ......
Page 90: ......
Page 156: ......
Page 184: ......
Page 225: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Page 253: ......
Page 254: ......