Section 5 —Preventative Maintenance
Page 5-10
VitalCare
TM
506N3 Series
Service Manual
Criticare Systems, Inc.
Electrical Safety Tests
Perform this test whenever the monitor housing is opened before
using the monitor on patients.
Withstanding Voltage
Test (Hi-Pot)
• High Voltage! The following test procedure requires working with
exposed electrical circuits and should only be attempted by
experienced technicians.
Equipment Needed
This test requires a Kikusui 7512DT Associated Research (or
equivalent) Hi-Pot tester. The tester should be rated for 4000 VAC
maximum.
Setup Hi-Pot Tester
• Hi limit = 1mA
• Lo Limit =0.0uA
• Arc Fail = OFF
• Arc Sense = 5
• Ramp Hi = Off
• Charge Lo = 0.0 uA
• Ramp Time = 1 second
• Dwell Time =1 second
• AC/DC =DC
WARNING
!
!