Section 5 —Preventative Maintenance
Page 5-18
VitalCare
TM
506N3 Series
Service Manual
Criticare Systems, Inc.
14.Access the
ALARM VOLUME
located in the
ALARMS MENU
.
Press the MENU key. Select
ALARM MENU
. Press the MENU
key again.
ALARM VOLUME <- 4
should appear. Press MENU
key again and the arrow turns (
ALARM VOLUME -> 4
). Verify an
audible change from 0 to 10 with each volume level setting.
15.Press the alarm SILENCE key. Verify that a red LED lights up
the alarm silence indicator on the front membrane. Press and
hold the alarm SILENCE key. Verify the red LED flashes
continuously and all alarms are silenced.
16.Restart the monitor to reset the NIBP module to
ON
.
Power Supply Performance
This applies to all 506N3 Series monitors.
1. Verify the green AC LED lights up the AC power symbol on the
front membrane when the monitor is plugged into the AC inlet.
2. Verify the monitor powers up on AC only.
3. Verify the monitor powers up on DC only with a battery.
Printer Performance
This applies to all 506N3 Series monitors.
1. Press the PRINT button when in the normal operating screen.
Observe that the monitor prints out the currently displayed
information.
NOTE:
For non-printer units, attached a test printer to the unit
and verify that the monitor prints out what is currently displayed
information.
Summary of Contents for 506DN3
Page 12: ......
Page 56: ......
Page 62: ......
Page 90: ......
Page 156: ......
Page 184: ......
Page 225: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Page 253: ......
Page 254: ......