Section 6 —Service Testing & Calibration
Setup
1. Turn off the power and disconnect the AC (Mains) power.
2. Place the monitor on a ESD protected workbench. Observe all
ESD protection procedures as described in “Electrostatic
Discharge Protection” in Section 7.
3. Connect the download cable to the DB-9 serial connection of
the monitor’s front bezel. Connect the other end to the DB-9
serial port of the computer.
4. Open the service tool on the computer. Select
Start > Programs
> CSI NIBP Service > NIBPSvc
(or click on the shortcut on the
PC desktop). Close the disclaimer dialogue box.
5. Select
COM1
for the port.
For laptops using a USB adapter
, select
COM 4
or an alternate
COM port as necessary.
6. Select
506N3 NIBP MODULE
in the drop-down menu box.
Figure 6-3: Select the COM Port and Model
7. Plug the monitor back to the AC (Mains) source.
8. Power up the unit while pressing the NIBP START/STAT/STOP
key.
9. On the display of the unit, verify that the message
CHECKING
FOR NIBP TOOLS….
appears.
10.Select the
Connection
from the drop-down menu. Select
Open
in Service Mode
.
Figure 6-4: Select Open in Service Mode
11.If communication has been established, the following screen
should appear. When this message appears, communication
with the PC is established.
Figure 6-5: Communication Established
5
6
Summary of Contents for 506DN3
Page 12: ......
Page 56: ......
Page 62: ......
Page 90: ......
Page 156: ......
Page 184: ......
Page 225: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Page 253: ......
Page 254: ......