Criticare Systems, Inc.
VitalCare
TM
506N3 Series
Service Manual
Page 6-9
Section 6 —Service Testing & Calibration
3. Click
Next
and verify that the pressure at the cuff is “0mmHg”
±2mmHg.
Figure 6-8: Verify Cuff Pressure
4. Connect the fitting from the manometer onto the NIBP fitting on
the unit. Click
Next
again. The pump should inflate to
250mmHG and then settle. Use the
UP
or
DOWN
buttons to
adjust the gain to match the manometer pressure with the cuff
pressure.
5. When Manometer and cuff match, select
Finish
.
Figure 6-9: Finish Calibration
6. The cal information is stored.
Figure 6-10: Storing Data