46
Troubleshooting
If the system stops working properly, check whether you are able to correct the
problem yourself based on the tips below. If this is not possible, contact the
customer service department.
Do not attempt to repair a defective electrical device yourself!
Problem
Possible cause
Fix
The cuffs do not
pump up within
15 minutes.
Air escapes between
the connection tubes
and tube connectors.
• Push the ends of the con-
nection tubes onto the tube
connectors of the cuffs and
control unit as far as they will
go.
• Check to see if the ends of
the connection tubes are
fi
rmly attached to the tube
connectors or whether they
have widened. If they have
widened, cut off the widened
ends.
The pump is not
activated.
Press the
button to pump up
the cuffs. Set the pressure regu-
lator to the highest level.
The air input is
obstructed.
Make sure the connection tubes
and cuffs are not bent or twisted.
The cuffs do
not de
fl
ate
completely.
The air extraction func-
tion is not activated.
Press the
button to extract air
from the cuffs. Set the pressure
regulator to the highest level.
The air
fl
ow is
obstructed.
Make sure the connection tubes
and cuffs are not bent or twisted.
Summary of Contents for 001-1011-00
Page 2: ......
Page 50: ...50...
Page 75: ...75 Service de distribution DS Produkte GmbH Stormarnring 14 22145 Stapelfeld Allemagne...
Page 76: ...76...
Page 101: ...101 Distributie DS Produkte GmbH Stormarnring 14 22145 Stapelfeld Duitsland...
Page 102: ...102...
Page 127: ...127 Sprzeda DS Produkte GmbH Stormarnring 14 22145 Stapelfeld Niemcy...
Page 128: ......