DT_0006-15
Malfunction
Cause of error
Rectification
is displayed on the monitor.
entered.
correct PIN “1111”.
The ECG device is not
switched on.
Switch the device on.
Bluetooth dongle is not plugged
into USB port on the PC.
Insert Bluetooth dongle in a free
USB port.
Device is paired with another
monitor.
Reset the device by pressing the
button for 20 seconds.
When the monitor is started,
the error “Bluetooth error”
No dongle…” occurs.
Bluetooth dongle is not plugged
into USB port on the PC.
Insert Bluetooth dongle into a free
USB port.
Bluetooth driver software is not
installed.
Install the Bluetooth driver software.
For this, follow the installation
instructions included in delivery.
The ECG device cannot be
found under the “Find
sensor” settings.
The ECG device is already
paired. In this state, the device
is not visible.
Reset the device by pressing the
button for 20 seconds.
During recording, black bars
are displayed in the ECG
signal once in a while.
There are transmission errors.
Bring the ECG device back into
range of the receiver. Check the
positioning and skin contact of the
electrodes.
The ECG device is out of
range.
One or several electrodes do
not have proper contact.
After the end of the
recording, black bars are
occasionally displayed in
the ECG signal.
There were transmission errors
during recording.
Check the positioning and skin
contact of the electrodes.
The ECG device was out of
range temporarily during
recording.
One or several electrodes did
not have proper contact during
recording.
During recording,
pacemaker pulses were
detected (incorrectly) and
marked in red.
When the electrode contact is
poor, it can lead to false
pacemaker detection.
Check the positioning and skin
contact of the electrodes or
deactivate the pacemaker detection
for subjects without pacemakers.
The ECG device cannot be
found after disconnecting
The Bluetooth settings are
initialised on application start-
Restart Spirotrac.