background image

 

Version 1.0A 

 

UM08B-CSZ064C 

User’s Guide

 

   

 

Summary of Contents for UM08B-CSZ064C

Page 1: ...Version 1 0A UM08B CSZ064C User s Guide ...

Page 2: ......

Page 3: ...ontents 2 1 2 B CSZ064C Lens Assembling 2 1 3 IR Remote controller 5 1 4 Microscope menu functions 8 1 5 Microscope introduction 22 1 6 Microscope Focus 26 1 7 Microscope accessory 28 2 Printed Notice 29 2 1 Maintenance 29 2 2 Product Specification 30 ...

Page 4: ...ble 1 IR Remote Control 1 Power adaptor 1 User s Guide 1 S114 Stand 1 2x B CSZ064C Lens 1 Stand assemble guide 1 1 2 CSZ064 Lens Assembling 1 2 1 Fix microscope Each CSZ064 lens has reach its parfocal distance when goods ship out simply screw a the CSZ064C lens under the microscope camera and tighten up with stand b refer to 1 5 3 section for assembling ...

Page 5: ...3 1 2 2 Start to use microscope Connect the HDMI cable c and adaptor d within the box to HDMI monitor and power See below a b ...

Page 6: ...4 1 2 3 Power On Off Press Power to turn on microscope Choose HDMI mode or Auto on HDMI monitor to output image c d ...

Page 7: ... when connect to optional accessory Image Capture Box IMB05 for snapshot 5 LED Adjustment No Function 6 Motor Reset Recalibrates entire operation Returns to the lens position If a loss of electricity occurred during operation please reset the operation 7 Manual focus Press Zoom out till the end to set up the sensor motor at the top so that all optical objectives lens reach the parfocal distance ...

Page 8: ...please refer to Section 1 6 Microscope Focus Press button one time to search focus in current area press and hold button for 2 seconds to search focus in whole area 8 1 Single Autofocus Focuses one time 8 2 Continuous Autofocus Re focuses when image blurred 9 Exposure Value 18 levels of exposure value are selectable from 9 Auto 9 9 1 Increase Use to increase exposure when the image is too dark 9 2...

Page 9: ...will show OZ sensor position and EV Exposure Each status only shows up 3 seconds 17 Auto White Balance Calibration To calibrate the white balance place the white balance card included in box under the lens focus until clear and choose On Because the white balance is sometimes influenced by external light sources it is recommended to calibrate the white balance if the color seems off or incorrect ...

Page 10: ... 4 Microscope menu functions There are 8 functions in menu press Up and Down to choose item Left and Right to turn on off or adjust parameter Enter means more functions press Menu to enter next function or confirm means no more settings Return to return last page or press Left or Right to choose End to return preview window ...

Page 11: ... Control auto control the brightness based on setting when light is not enough 3 SENS UP Only available when item 2 AGC value is larger than 0 use to increase Senor to light s sensibility 4 Brightness Adjust image brightness 5 D WDR Digital wide dynamic range Digitally adjusting exposure in areas of the frame to maintain optimum detail in both the shadows and highlights of the image ...

Page 12: ...near the edge of defog area 6 3 DEFAULT Return to default setting 6 4 RETURN Return to last page 1 4 2 BACKLIGHT Press left right button to choose turn on off Backlight mode 1 OFF Disable backlight function 2 WDR Wide Dynamic Range WDR allows an imaging system to correct for the intense back light surrounding a subject and thus enhances the ability to distinguish features and shapes on the subject...

Page 13: ...to last page 3 BLC Back Light Compensation To adjust brightness for darker image in back light environment 3 1 GAIN To adjust brightness level 3 2 AREA Choose the back light compensation area and size 3 3 DEFAULT Return to default setting 3 4 Return Return to last page or press left or right to choose End to return preview window ...

Page 14: ...djusted 4 areas are selectable 4 2 DISPLAY Display the selected area 4 3 LEVEL Set value to judge if the image is too bright 4 4 MODE Super back light compensation mode 4 5 ALL DAY Day mode 4 6 NIGHT Night mode 4 7 BLACK MASK Reverse the over exposed area 4 8 DEFAULT Return to factory default setting 4 9 RETURN Return to last page 1 4 3 MANUAL WB Calibrate the white balance ...

Page 15: ...ack the white color from current image 4 MANUAL Manually adjust blue and red values to find correct white color Refer to below fig 1 4 4 NR Noise Reduction NR To obtain high quality output image and increase file compression See below 1 2D NR ON OFF Edge Preserving 2D NR 2 3D NR ON OFF Motion Adaptive 3D NR See below ...

Page 16: ...er value gets higher sensitivity 3 1 2 RETURN Return to last page 3 2 LEVEL The threshold to judge noise 3 3 START AGC Start auto gain control When AGC value is higher than setting will start auto gain 3 4 END AGC End auto gain control When AGC value is higher than setting will end up auto gain 3 5 Return Return to last page 3 Return Return to last page ...

Page 17: ... few more functions in D EFFECT but 1 3 1 4 and 1 6 function can only turn on one item at one time Press Left or right to turn on or off See below functions 1 1 FREEZE Turn ON OFF to freeze image If Freeze function is turn on the 1 2 Mirror function can not be ...

Page 18: ... D ZOOM Digitally enlarge the chose area Digital zoom selection can from 1 62 times See below setting for D Zoom 1 3 1 PIP Turn on Picture in Picture function When enable PIP function the right bottom corner will show the original image The rest of image is D zoom enlarge image ...

Page 19: ... DEFAULT Return to default setting 1 3 5 RETURN Return to last page 1 4 SMART D ZOOM Assign 1 or 2 areas to enlarge the image from 1 62 times whenever detected the object moves The Smart D zoom will automatically enlarge the assigned area for few seconds and then return to original image ...

Page 20: ...area can be set in 1 4 1 1 4 3 SENSITIVITY The sensitivity to detect whenever object moves The lower value gets the higher sensitivity 1 4 4 PIP No Function 1 4 5 D Zoom Set magnification times 1 4 6 TIME Set the enlargement waiting time 1 4 7 DEFAULT Return to default setting 1 4 8 Return Return to last page ...

Page 21: ... Select the areas which needs to be detected 4 areas are selectable 2 2 DISPLAY Display the selected area Position and area can be set in 2 1 2 3 SENSITIVITY The sensitivity to detect whenever object moves The lower value gets the higher sensitivity 2 4 MOTION VIEW Mark specific area to identify the moving object 2 5 DEFAULT Return to default setting 2 6 RETURN Return to last page ...

Page 22: ...d Pixel Correction Correct the live dead pixel during preview 4 2 1 LEVEL Set the threshold value to judge if it needs RETURN Return to last page 4 2 STATIC DPC Static Dead Pixel Correction scan and find the dead pixel then save the position for correction Note Please do not change this setting ...

Page 23: ... the dead pixel 4 2 4 AGC Auto Gain Control When light is not enough AGC will follow the threshold setting to compensate the dead pixel 4 2 5 RETURN Return to last page 4 2 2 to be corrected 5 VERSION Firmware version 6 RETURN Return to last page 1 4 6 ADJUST See image adjust as below fig 1 SHARPNESS Adjust Sharpness higher value increases the higher contrast along near edges of the image ...

Page 24: ...actory default setting Press Left right button to choose CHANGE to return default 1 4 8 RETURN Exit from menu 1 5 Microscope introduction 1 5 1 IR Remote Area Indicator When both f and g LED indicator blinks the microscope is in standby mode Caution Please remove the transparent protection sheet on the IR receiver area ...

Page 25: ... blink whenever buttons are pressed If the indicator did not blink means the signal was not delivered and please repress again 2 HDMI indicator f When lit up the microscope is in HDMI mode If light blinks means microscope does not connect with monitor f e g ...

Page 26: ...sed to connect the microscope to a screen When the device is in HDMI mode connect the microscope to the screen using the HDMI cable from the box For TV screen connections make sure to choose the homologous image ratio and format 2 USB i USB port to use optional accessory IMB04 to PC 3 Power j Power adaptor is only compatible with the cord in the box h i j ...

Page 27: ...tor Assembling Assemble the stand first Insert CSZ064C lens into Ring Adaptor k and screw tight with Hex Key Insert lens with Ring adaptor into Universal Adaptor then screw tight l the lens face front center position k l ...

Page 28: ...he coaxial light m to lens then switch power ON n 1 6 Microscope Focus Move the B CSZ064C lens upward leave 35mm from the object press Zoom out button move the sensor to end position Turn the optical zoom to 4x o move up and down the universal gear box p to find the clear image m n ...

Page 29: ...e is clear turn the optical zoom from 4x to 0 6x to see if image is clear at each magnification If clear at each stage the focus is done If the image is not clear at each magnification please re try from beginning p o ...

Page 30: ...ens working distance 35mm 1 38 2X CSZ064C Lens magnification on 24 HDMI monitor Ref Data Optical Adaptor 1x 0 6x 1x 2x 3x 4x 2x Object Lens Working Distance mm 35 35 35 35 35 Field of View mm 4 65 2 9 1 4 0 95 0 7 Magnification X 101x 163x 380x 500x 650x 1 7 Microscope accessory 1 7 1 White balance card The white side is used to calibrate the white balance If the object is tiny place it onto the w...

Page 31: ...alfunction if it encounters strong collision vibration or distortion 2 1 4 Turn the microscope off before cutting off the power supply do not forced cut off the power supply 2 1 5 Do not face the lens against strong light or sunshine for extended periods of time Strong light rays may degrade sensitive elements and generate white stains on images 2 1 6 Handle the device carefully do not disassemble...

Page 32: ... DC port Adaptor DC Input HDMI port Output 1080P 1920 1080 Pixel image Focus control IR remote control Power supplier Adaptor Output DC 5 0V 2 0A Input AC 100 240V 50 60Hz Power Consumption A C 0 225A Max Size 280 L 66 W 68 T mm Weight Host weight around 580 grams Operation environment Temperature 5 50 Humidity lower than 85 No Condensation For any changes please visit http www vitiny com ...

Page 33: ... the device or the device is dropped or damaged please switch off and remove the power supply to avoid fire and electric shock Do not look steadily into Light sources after switching on the host as it is harmful to your eyes Please use only the product accessories provided in the box for connections Do not use any outside products without approval of original factory Please remove the power supply...

Page 34: ...er etc Services caused by any of the following situations in the period of the warranty shall not be provided free of charge 1 Improper use or disassembly repair or refitting 2 Any damages to the device caused by external or environmental factors 3 Discrepancy of product serial number unfilled or unidentified warranty Any inspection or repair service after the period of warranty will be charged as...

Page 35: ...l for Confirmation Stamp is necessary for validation of the Warranty Distributor s seal shall include name of the shop telephone and address Please ask the distributor to fill in the name of the shop address purchase date and other contents to protect your rights and validate your one year warranty since upon the purchase date ...

Page 36: ...HTTP WWW VITINY COM MicroLinks Technology Corp All rights reserved ...

Reviews: