Instructions for Use
23
6.3.3
Using PCR Plates in the PCR Adaptor Block (Type 1)
1. Insert the semi-skirted or un-skirted PCR plate into 96 well adaptor block
(see Table 7), making sure that it is level with the top of the block and that
it is securely and fully inserted.
2. Insert the adaptor block into the plate holder on the unit.
6.3.4
Using Micro Tubes in the Tube Adaptor Blocks (Type 1, 2 and 3)
1. Select the appropriate adaptor block for the tube being used from Table 7.
2. Insert the micro test tubes into the adaptor block, ensuring that they are
securely and fully pushed down in the wells.
3. Insert the adaptor block into the plate holder on the unit.
4. After mixing, remove the adaptor block from the unit and then remove the
micro tubes as required.
6.3.5
User-defined Mixing Programs and Profiles
The unit can store over 90 user-defined mixing programs, which can be
created by copying an existing program and then changing the program ID
and the speed and time parameters as required. See section 7.1.2 for
details.
User-defined mixing programs can also consist of a series of individually
defined program steps, each with different mixing parameters and run in
succession to produce a complex profiled mixing sequence. Refer to section
7.2 for details.