earth ground continuity the Vitrek 930i Ground Integrity Analyzer is recommended.
To start low resistance testing, press the low ohms key. The low ohm prompt will appear as follows:
LOW OHM TEST
DIRECT
AUTOMATIC
To conduct manual low resistance measurements, select the direct option. The 944i uses a 2-wire measurement technique.
Proper zeroing of test lead resistance is essential to obtaining reliable results. To facilitate the canceling of lead wire
resistance the 944i will provide a brief message as follows:
PRESS 0 TO NULL
LEAD RESISTANCE
It is recommended that the user "null" the lead set periodically to obtain the most accurate results. Always do a "null" when a
new or different set of test leads are used with the unit. If you wish to null the test leads press the zero (0) key. When the "0"
is pressed the following prompt will appear:
SHORT V LO & CHASSIS
SENSE LEADS (ENTER)