DC-7 7kV DC output
Increases DC output range to 7kV DC (with decreased current limits). DC-7 option does not affect AC output specs. No
external boxes or circuitry are necessary.
DC-10 10kV DC output
Increases DC output range to 10kV DC (with decreased current limits). DC-10 option does not affect AC output specs.
DC-10 requires an external box with circuitry inside.
AC-8 8kV AC output
Increases AC output range to 8kV DC (with decreased current limits). AC-8 option does not affect DC output specs.
AC-8 requires an external box with circuitry inside.
FR-4 Fixed Range power supply mod
Option improves stability of unit’s output in special cases. Current output is limited in some voltage ranges. Option may
be installed by factory without being requested. The current limit specifications with the FR-4 option installed differ
from a standard unit.
RPO Rear Panel Outputs
Provides V-Hi, V-Lo, and Chassis Sense outputs at the rear of the unit.
FS-1 Four Wire Fail Safe option
Provides four wire Lo-Ohms measurement capability (V-Lo and Chassis sense outputs are BNC connectors) and limits
the current output in Lo-Ohms mode to 8.5mA.
PLC-2A Programmable Logic Controller interface
Option provides a 5-pin DIN connection in the rear of the unit for the purpose of controlling and monitoring automated
testing with an external PLC system.
ADC Automatic Discharge option
Option provides rapid discharge of the output. No external devices necessary.
LVIR Low Voltage Insulation Resistance option
Option provides the additional ability to measure Insulation Resistance from 10 to 99 VDC in IR mode. No external
devices necessary.
LO-500 Low Ohms 500 option
Option provides the additional ability to measure up to 500 ohms using the LO-OHMS mode. No external devices
necessary.
Routine Inspection and Maintenance