The 944i is equipped with a fan to cool the internal electronic circuitry. Periodically it is advisable to inspect the filter
assembly located on the rear panel. In areas where there is a high level of dust or other airborne particulate matter, the
inspection of this filter should be conducted more often. If the filter is allowed to become restricted, overheating can
occur causing damage to the electronic components.
Should the filter require cleaning, the finger guard snaps off to provide easy access to the filter element. A thorough
rinsing in a warm soap water solution should remove any build-up. Insert the filter element back into the finger guard
and snap into place. Note that the finger guard has four tabs which should line up with the four indentions on the fan
mounting bracket.
.........................................................................................